| Location | Ultramicroscopy Hall (GW0.016) |
|---|---|
| Functions | SEM, EDX, GIS, nanomanipulator |
| Specimen | About 20 samples that fit on a 12mm diam. stub, beware that if you would like to image high samples, that you should discuss it with the responsible technician. |
| Manufacturer | ThermoFisher |
| Useful files |
- SEM manual - Paper 'Trinity detectors' - Periodic table for EDX |
| Responsible | Luc Wigbout |
The ThermoFisher Apreo SEM is a Scanning Electron Microscope with a standard Everhart Thornley Detector (ETD) and three in-column "Trinity" detectors. The Apreo SEM has some additional detectors, enabling several other probing techniques in-situ, e.g. energy dispersive x-ray spectroscopy (EDX) to quantitatively measure spectra and qualitatively create spectral maps at the micro- and nanoscale.
Large (tall) samples should be discussed first with the responsible technician!
If the Sample Exchange window is not showing, press the
-button in the upper-left.
If the live CCD image is not showing, first close the Sample Exchange window, and check the lower-right window of the quad-view. Make sure that the right detector-type is selected. There should be a light-bulb in the lower-left corner.
When you walk in, you should see the Sample Exchange window, which should look something like the figure on the right. The grayscale image is a live CCD image of the chamber, where you can see the stage at the bottom, the conical objective at the top, and the GIS needles at an angle.
The screen on the left should have an Excel sheet open, with the logs from previous users. Check if there are any comments from previous users that are important to keep in mind during your session!
Before venting the system, check the chamber pressure on the right screen in the bottom left. The previous user should have pumped the chamber to high vacuum again, so it should be <1e-5 mbar. Write the pressure in the logbook! Furthermore, write down the emission current and the electron source pressure!
If not done already, mount your sample on a SEM-stub, using either carbon- or coppertape, or carbonpaint.
Open the chamber while looking at the live CCD image, to make sure you do not touch the objective lens.
Load your sample onto the stage. Of course, you do this while wearing orange (powder-free) gloves! When closing the door, make sure that your sample does not touch (and thus damage) the objective lens!
In the Sample Exchange window, make sure that 'High Vacuum' is selected.
Make sure to check the box 'Take Nav-Cam Photo', so that the stage will automatically move into the Nav-Cam position and take an image.
If you have a magnetic sample: check the box of 'Magnetic Sample'.
Pump the chamber, choose 'no accessory'.
While the chamber is pumping, you want to change the subfolder to D:\User\<yourname> on the Support-PC. and close the Sample Exchange window.
Check if the settings are in default:
Check for the top two quad-view windows:
Check additional settings:
Set the dwell time to 100 ns, this is recommended during imaging, as the refresh rate will be quick enough to see the effects of turning the knobs (focus, stigmator, etc.) quickly.
Set the resolution to 3072x2048 or smaller. Typically, high resolutions are not necessary during imaging and make the refresh rate unnecessarily slow. Taking HQ micrographs, however, is useful sometimes.
If the chamber pressure < 9.9e-5 mbar you can turn the beam on in the Beam Window.
On the Nav-cam, double-click the highest sample and hold your hand above the ESC-button.
Everytime you move the stage, you should hold your hand above the ESC-button so that you can abort movement immediately when the stage touches anything!
Obtain a clear image of the sample: zoom -> focus -> zoom -> focus -> etc.
Get a good focus at >20 000x magnification, so that the calculated Working Distance (WD) has a small error. (While rotating the focus knob, notice that the WD in the blue bar of the quad-view changes.)
Open the Stage window, and notice the red arrow pointing up, denoting that the stage is not linked. Also notice the Z-height. This is defined from the bottom of the chamber to the top.
Press 'Link Z to FWD'-button
, which then changes to
.
Again, look in the Stage window, and notice how the arrow is switched to a white arrow pointing down
. Also notice that the Z-height changed its value, as the Z-axis is now defined from the objective lens downward!
In the Stage window, move the Stage to Z = 10 mm. This will leave 10 mm between the objective lens and the sample, i.e. bring the sample up toward the yellow line you see in the live CCD image.
At Z = 10 mm, refocus on your sample.
Relink the FWD, and move again to 10 mm. Now the working distance should equal the Z-distance.
Gather data!
When you're done, make sure to leave the Apreo behind properly.
Your session is coming to an end. 15 minutes in advance, you will have to close down by following these steps:
D:\User\<yourname> on the Support-PC to a USB stick or to \\data03.physics.leidenuniv.nl\microscope\<yourname> if necessary.No USB-sticks are allowed in the Microscope PC (the PC on the right)!
This section describes some useful tools during imaging. To speed up your workflow, you could check out the Keyboard Shortcuts in the manual or click on the image on the right.
Remember SEM stands for scanning electron microscope, this means that the electron beam is scanning across your sample at high speeds due to a potential change in the x- and y-deflectors.
On the left of the quad-view image, you see an orange line which shows you where on the slow-axis (vertical scanning) it is scanning.
Since the fast-axis (horizontal scanning) is so quick, there is nothing that depicts that axis.
So, the beam is scanning across your sample and the detector is simply measuring the amount of electrons 'reflected' from that part of the surface.
The detector measurement and the x/y-deflection of the beam are phase-locked, such that the image is generated.
Remember that there is no live image with this technique.
In order to properly focus on your sample, it is useful to have a feature in your image with some contrast.
To get some contrasting features, change the contrast and the brightness; these act as gain and offset respectively.
These gain and offset are actual physical parameters (voltages) that you change in the detector; these are not the digital after-the-fact changes to the image alone!
After you actually get an image with some information, you are ready to focus onto a small region of your sample.
At some point, focusing will come to a halt, as you will see that only parts of your sample will be in focus, while other parts are not.
In order to continue, try to obtain a focus such that all parts are 'equally blurry', i.e. 'equally in focus'; this is called the plane of least confusion.
It is easiest to find this plane using features that have some contrast in X- and in Y- direction, as to check the blurriness of that contrasting part.
IMAGE IMAGE IMAGE IMAGE IMAGE IMAGE
If the stigmator is not working as you think, it might be useful to check in the Beam Settings Window if the small stigmator square is somewhere around the middle.
If the stigmator is (still) not working as you think, check if you are actually in the plane of least confusion, you probably aren't.
If the sample is shifting in X- and Y-direction during the rotation of the stigmator knobs, you might want to check out the stigmator alignment under the Direct Adjustments tab.
XYZ and R,T.
Stage > Align Feature
SNR increases in time and electron counts.
Resolution / Dwell Time / Reduced Area
20us no integration
However: driftless settings 1 us, 20x integration and drift correction
White paper
Lens alignments
Cross section
More tips and tricks will come... If you have any tips, we encourage you to contribute to the NanoWiki!